TH
-130 |
| |
 |
- Impact Device D integrated: no cables!
-
Wide
measuring range in HLD and
direct display of converted
hardness values in HB, HRB, HRC,
HRA, HV, HS
- For
most metals
- Test
at any angle
- Simple
handling and low test
expenditure
- Optional
printer TA220S available
|
|
|
|
|
Measuring Range |
| Material |
HLD |
HRB |
HRC |
HRA |
HB |
HV |
HS |
| Steel
& cast steel |
300-900 |
38.4-99.8 |
20-38.4 |
59.1-85.8 |
81-854 |
81.1-955 |
32.5-99.5 |
| CWT.ST |
300-840 |
|
20.4-67.1 |
|
|
80-898 |
|
| Stainless
steel |
300-800 |
46.5-101.7 |
19.6-62.4 |
|
85-655 |
85-802 |
|
| GC.
Iron |
360-650 |
|
|
|
93-334 |
|
|
| NC.
Iron |
400-660 |
|
|
|
93-334 |
|
|
| C.
Alum |
200-570 |
23.8-34.6 |
|
|
26.8-164 |
|
|
| Brass |
200-550 |
13.5-95.3 |
|
|
40-173 |
|
|
| Bronze |
300-700 |
|
|
|
60-290 |
|
|
| Copper |
200-690 |
|
|
|
45-315 |
|
|
|
|
|
|
Technical
Specifications |
|
Standard
Impact Device |
D
Integrated |
|
Hardness
Scale |
HLD,
HB, HRC, HRB, HV, HS |
|
Measuring
range. materials |
See
table above |
|
Accuracy |
+
6HLDL |
|
Memory |
99
average readings |
|
Output |
RS
232 to printer |
|
Min.
Surface Roughness of Workpiece |
1.6
um (Ra) |
|
Max.
Workpiece hardness |
900HLD |
|
Min.
radius of Workpiece |
Rmin
= 50mm |
| Min.
Workpiece weight |
2-5
kg on stable support 0.05-2kg with compact
coupling |
|
Min.
Workpiece thickness coupled |
5mm |
|
Min.
Thickness of hardened layers |
0.8
mm |
|
Indentation
depth |
Impact
Device data |
|
Continuous
working time |
8h |
|
Power |
Rechargeable
Li-Polymer battery |
|
Operating
temperature |
0-40°C |
|
Overall
dimensions |
155
x 24 x 55mm |
|
Weight |
180
g |
|
|
|
|
|
Standard
delivery |
- Main
Unit integrated with Impact
Device D
- Test
block with HLD value
- Charger
- Cleaning
brush
- Time
certificate
- Instruction
manual
- Carrying
case
|
|
1
1
1
1
1
1
1 |
|
|
|
Optional
accessories |
- Support Rings
- Printer
TA220S with cable
|
|
|
|
|
|
| |
|
|
|
|
|